GO2 Oxygen saturation meter

GO2 Pulse Oximeter

Oxygen saturation meter

Najít podobné výrobky

The accurate, affordable and durable GO2 pulse oxymeter allows your patients the freedom to go about their activities, while delivering accuracy comparable to professional oximeters.

Specifications

General system
General system
Oxygen saturation display range
  • 0 to 100 %
Pulse rate display range
  • 18 to 321 beats per minute beats per minute
Oxygen saturation declared accuracy range (Arms*)
  • 70% to 100% SpO2 ± 2 digits %SpO2
Pulse rate declared accuracy range (Arms*)
  • 20 to 250 BPM ± 3 digits beats per minute
Low perfusion pulse rate declared accuracy range (Arms*)
  • 20 to 240 BPM ± 3 digits beats per minute
Measurement wavelengths and output power
Measurement wavelengths and output power
Red
  • 660 nanometers @ 0.8mW max average nm
Infrared
  • 910 nanometers @ 1.2mW max average nm
Temperature
Temperature
Operating temperature
  • 5°C to 40°C* (41°F to 104°F)
Storage/transportation temperature
  • -30°C to 70°C (-22°F to 158°F)
Humidity
Humidity
Operating humidity
  • 10% to 90% relative humidity, non-condensing %
Storage/transportation humidity
  • 10% to 90% relative humidity, non-condensing %
Operating altitude
  • Up to 4,000 meters/13,000 feet
Battery life
Battery life
Continuous
  • Approximately 2400 spot checks based on approximately 21 hours of operation using one AAA size alkaline battery, calculated at 20 seconds per use.
Storage
  • 6 months minmum
  • Device temperature will not exceed 41°C as measured during a controlled environment test.