Nové
Tempus IC2 Integrated monitor and telemedicine solution

Tempus IC2

Integrated monitor and telemedicine solution

00-2001

Najít podobné výrobky

An advanced portable multi-parameter vital signs monitor with integrated telemedicine. Tempus IC2 allows trained non-medical users to manage medical emergencies in the air, at sea, and in remote locations. Along with built-in ease of use, the Tempus IC2 features clear visual guidance and automated data collection, helping users to feel confident in managing a medical emergency.

Specifikace

Physical dimensions
Physical dimensions
Standalone size
  • 263mm x 216mm x 98mm (10.4" x 8.5" x 3.9")
Standalone weight
  • 3 kg (6.6 lbs) including battery, RapidPakᵀᴹ clip, NBP cuff, hose, communications cable and headset
Bag size
  • 401mm x 276mm x 154mm (15.8" x 10.9" x 6.1")
Weight when packed in bag
  • 5.9 kg (13 lb) Tempus IC2 plus accessories (excluding power supply, and mains lead/battery charger)
Thermometer
Thermometer
Type
  • Handheld, Bluetooth-enabled infrared tympanic temperature
Displayed range
  • 32–43°C (89.6–109.4°F) (target)
Display resolution
  • 0.1°C (0.1°F)
Accuracy
  • ±0.2ºC (36–39°C) ±0.4°C (<lt/>36 <gt/>39°C) [±0.4ºF (96.8-102.2°F) ±0.5ºF (<lt/>96.8, <gt/>102.2°F)]
Bluetooth communications range
  • 0–10 m (0–32'9")
12-lead ECG
12-lead ECG
Heart rate range
  • 30–300 bpm
Input impedance
  • <gt/>100 MΩ
Dynamic range
  • ±5 mV ac
Accuracy
  • ±3 %
DC offset
  • ±300 mV dc
Frequency response
  • 0.05Hz–175Hz ±3 dB⁵
Sample rate
  • 500 Hz
Common mode rejection
  • 95 dB min, additional filters include mains, muscle, low and high pass and adaptive baseline zeroing
Protection
  • Protected against defibrillator charge at 5 kV
Capnometer - respiration rate
Capnometer - respiration rate
Range
  • 0–149 bpm
Accuracy
  • 0–70 bpm: ±1 bpm / 71–121 bpm: ±2 bpm / 122–149 bpm: ±3 bpm
Glucometer
Glucometer
Test range (preset to mg/dL)
  • 10–600 mg/dL (0.6–33.3 mmol/L)
Reading time
  • 3 seconds
Bluetooth communications range
  • Class 2, range 0–3 m (0–9' 10") in an open field
Pulse oximetry - pulse rate
Pulse oximetry - pulse rate
Range
  • 25-239 bpm
Accuracy (all ages)
  • No motion ≤3 digits; motion ≤5 digits
Pulse oximetry - SpO₂
Pulse oximetry - SpO₂
Range
  • 1–100 %
Accuracy (adult/child)
  • No motion/low perfusion: ±2 digits 70–100%; motion: ±3 digits, 70–100%
Perfusion index range
  • 0.02–20%
Noninvasive blood pressure (NBP)
Noninvasive blood pressure (NBP)
Adult/large cuff range
  • 20–260 mmHg
Adult/large cuff accuracy
  • ±3 mmHg or ±2% (whichever is greater)
Child cuff range
  • 20–260 mmHg
Child cuff accuracy
  • ±3 mmHg or ±2% (whichever is greater)
Battery and power
Battery and power
Charger mains input voltage
  • 100–240 V
Charger frequency
  • 50–60 Hz & 400 Hz
Charger input currency
  • 0.9 A max at approximately 100 V
Charger output voltage
  • 8.4 V dc
Charger output current
  • <lt/> 2.73 A
Charger weight
  • 0.25 kg nominal
Charger dimensions
  • 107mm x 67mm x 36.5mm (4.2" x 2.6" x 1.4")
Power supply, external⁴
  • PSU-rated 100–250 V, 50 Hz, 115 V 400 Hz
Battery type
  • Single, external, user-replaceable, lithium-ion
Battery strength indicator
  • Integral
Battery life
  • Min 11 hours¹ [default display brightness, SpO₂/EtCO₂ (25% of the time) and NBP every 15 minutes]
Battery capacity
  • Nominal 7.4 V, 10.2 Ah
Battery shelf life
  • Approximately 7 hours remaining after 1 year storage²
Battery charge time from empty
  • 6³ hours
Integral digital camera
Integral digital camera
Annotations
  • Images can be annotated with text, colors, shapes which can be sent to Tempus IC2 by the hospital
  • ¹ Battery shelf life and run times are based on a new, fully charged battery stored and used at 20°C. Run time is based on RDT’s model of typical device usage in an incident.
  • ² Subject to conditions of storage and use.
  • ³ Only the RDT Battery Charger (part number: 01-1012) can be used with the Tempus IC2.
  • ⁴ Battery may be charged (optionally) by the Tempus IC2 when running on mains power. Internal charging can be restricted if required.
  • ⁵ Note that during monitoring prior and post recording ECG frequency response filters will be 0.5–175 Hz.